Activ W510 Aw-meter

  • Fanless chilled mirror technology
  • Robust, fast, accurate
  • Very easy cleaning, limited fouling
  • User-friendly interface
  • Automated traceability with Linko PC sofware
  • Low operating costs

 

 

Activ W510 specifications
Technology used Chilled mirror, single-stage thermoelectric module
Measurement range Aw 0.030 à 1.000 (HRE: 0.5 à 100%)
Accuracy Aw* +/- 0.003 (HRE: +/- 0.3%)
Repeatability Aw* Standard mode : ± 0.001 (HRE: +/- 0.1%)
Precision mode : ± 0.0005 (HRE: +/- 0.05%)
Room temperature range 0°C à 50°C
Auto detect Water or solvent condensation
Sensor soiling detection
Non-volatile memory 5000 measurements
1024 product datasheets
200 users, 10 users parameters banks
Number of measurement modes Variable time : 3 (Fast, Standard, Precise)
Fixed time : 6 (Flash, 2 min … 1 hour)
Isotherms mode : 2 (consecutive temperature plateaus)
Series of measures One shot to 250 successive measurements
Measurement time Fast mode : 2 to 5 min, depending on your product
Standard mode : 3 to 10 min, depending on your product
Calibration Automatic with 14 different types of saturated / insatured salt solution or manually
A return to factory settings is possible
Dimensions W 24 cm , L 36 cm , H 18 cm
Weight 7.3 Kg without external power supply
Power consumption Active thermoregulation : < 75W/h
Inactive thermoregulation : < 15W/h
Standby : approximately 5W/h
Power requierement External, input 100-240Vac, output 12Vdc 6,25A
Consumables provided Set of 30 plastic dishes with lids
100 ml bottle of saturated salt solution
Warranty 3 years parts and labour

*control salts or homogeneous products, maintenance carried out according to recommendations